Single and Three Phase Power Analysers Model 4106A Series

Features

  • Suitable for frequency inverter drives
  • Large and bright display for up to 10 values
  • Scope function, 20 setups can be stored
  • Harmonics 1-99, Bar charts
  • DC-300kHz, 15mA-40A, 0.3V-1000V
  • 0.1% accuracy

For more information about this product contact Tinsley

 

Single and Three phase Power Analysers Model 4106A Series

Introduction

Tinsley offer two state of the art high performance power analysers, single and three phase. Unlike other instruments at this price level the 4106A is designed to cope with the extreme signals generated on frequency inverter drives. You don't have to worry about the signal wave forms. The analysers will always provide precise and reliable measurements. The large and very bright monitor lets you read the display values from a distance of up to 4m.

Simple to use

The 4106A can be used for something as simple as measuring the power requirement of your coffee machine to a task as complex as determining the pertinent power parameters of a frequency inverter driven system.
All values are displayed in large characters, easily readable even in darkened environments. The user menu makes operation easy. The measured values can either be printed, sent to a PC via IEEE- or RS232 interface, or to a chart recorder via the analog outputs. You can have all available options installed or just the ones you require.

Extraordinary features, attractive display

Much effort has been put into the design of the 4106A Power Analysers to give you the highest performance at low cost. The analyser inputs are all galvanically isolated, are broad band DC-300kHz, have a wide input range (0.3V - 1000V, 15mA - 40A) and have an exceptional common mode rejection for use in frequency inverter driven systems. The accuracy is 0.1% (0.05% versions are available). The bright LCD monitor displays up to 10 measured values in highly legible 9mm high numbers.

The Three Phase Power Analyser puts up to 32 measured values on the screen. You have the choice to display wave forms, bar graphs or trend plots. A unique feature of these instruments allows a combination of meter mode and graphic mode.

Designed to last

With future development and as features are added to the 4106A you will be able to obtain upgrades, at modest cost, through your local Tinsley distributor. The Power Analysers are warranted for one year and are supplied with a traceable calibration certificate.

Model 4106A Specifications

See the table below for a run-down of this product's specifications.

Voltage
8 ranges, 1-3-10-sequence; 0.3V, 1V, 3V, 10V, 30V, 100V, 300V, 1000V
 
Frequency range
DC, 0.1Hz-300kHz
 
Crest Factor
4:1 at 50% full scale (fs)
 
Input Impedance
>1MΩ
 
Common Mode 50Hz/100kHz
155dB/95dB
 
Standard accuracy 23°C ±3°C; rms, mean, rectified mean; for 0.3, 1V range, input >50% fs
improved accuracy
 
1Hz-1kHz
±(0.1% range + 0.1% range)
±(0.05% rdg + 0.05% range )
 
DC, 1kHz-10kHz
±(0.2% range + 0.2% range)
 
10kHz-100kHz
±(0.3% range + 0.04%/kHz range)*
* 0.3V range typical
 
100kHz-300kHz
±(0.3% range + 0.04%/kHz range), typical
   
Current
10 ranges 1-3-10-sequence; 15mA,50mA,150mA,500mA,1.5A,5A; 1,3,10,30,100,300A
 
Frequency range
DC, 0.1Hz-300kHz
 
Crest Factor
4:1 at 50% full scale (fs)
 
Common Mode
50Hz/100kHz
160dB/120dB
 
Standard accuracy 23°C
±3°C 5A-/Shunt input1
30A input1
improved accuracy 1Hz-400Hz±(0.05% rdg + 0.05% range)
 
1Hz-1kHz
±(0.1% range + 0.1% range)
±(0.1% range + 0.1% range)
 
DC, 1kHz-10kHz
±(0.2% range + 0.2% range)
±(0.9% range + 0.2% range)
 
10kHz-100kHz
±(0.3% range + 0.04%/kHz range)
±(0.3% range + 0.5%/kHz rdg)
1for 2 lowest ranges, input>50% fs
 
100kHz-300kHz
±(0.3% range + 0.04%/kHz range), typical
 
   
Power
80 ranges corresponding to the product V x A

 

Frequency range
DC, 0.1Hz-300kHz

 

Accuracy 23°C ±3°C

 

1Hz-1kHz
(Add accuracy percentage figures of current and voltage input and add 0.04%/kHz of Vrms x Arms / PF)
PF = 0 to ±1

 

DC, 1kHz-10kHz
PF = 0 to ±1

 

10kHz-100kHz
PF = 1
   
Frequency
0.1Hz-300kHz, A or V triggered; Accuracy ±0.1%
     
Computed Values
Accuracy; Reactive Power, Var = ±(VA2– W2)1/2; Apparent Power: VA = Arms Vrms; Power Factor: PF = W/VA; Crest Factor: CF = Ap/Arms, Vp/Vrms; Form Factor:FF = At/Arms, Vt/Vrms; Impedance: Z = Vrms/Arms; Total Harm Dist: THD = (Irms2 - IFund2)1/2/Irms
Add accuracy percentage figs of values involved incomputation.
   
Integrator
Energy, Charge; Accuracy Wh, VAh, Varh, Ah; Basic accuracy of integrated quantity.
 
   
Harmonic Analysis
Frequency range of fundamental
2.5Hz-100kHz
 
Range of harmonic
1-99
 
Accuracy, Harmonic current and voltage
 
2Hz-1kHz
±(0.2% rdg + 0.1% range)
 
1kHz-10kHz
±(0.5% rdg + 0.5% range)
 
10kHz-100kHz
±(0.7% range + 0.1%/kHz rdg),typical
   
Display
Blue liquid crystal graphic display with FL backlight; 64x120mm; 128x240 pixels
     
Power
AC, 50-400Hz; Fuse: Power
85V-240V; 2AF/30VA

   
Dielectric
Strength
Inputs to case or power supply
2.5kV/50Hz/1 minute
 
Line input to case
1.5kV/50Hz/1 minute
 
Input to Input
4kV/50Hz/1 minute
     
Dimensions
H x W x D; Weight
150 x 235 x 320mm; 4kg

 
Options
IEEE-488-2, RS232, Centronics printer output
 
4 Analog outputs, Output impedance 100Ω
0 - ±5V
 
4 Analog inputs, low range, input impedance 200kΩ
0 - ±1V
 
4 Analog inputs, high range, input impedance 200kΩ
0 - ±10V
 
Rack mounting kit
 
Humidity: KYG according to DIN 40040, max. 85% RH non-condensing
   

Versions

4106A-1/0.1
Single Phase Power Analyser, 0.1% accuracy, nonvolatile memory to save 20 instrument settings.
4106A-1/0.05
Single Phase Power Analyser, 0.05% accuracy, nonvolatile memory to save 20 instrument settings.
4106A-1/nh
Single Phase Power Analyser, without harmonics, without graphics, 0.1% accuracy, nonvolatile memory to save 20 instrument settings.
4106A-3/0.1
Three Phase Power Analyser, 0.1% accuracy, nonvolatile memory to save 20 instrument settings.
4106A-3/0.05
Three Phase Power Analyser, 0.05% accuracy, nonvolatile memory to save 20 instrument settings.
4106A-3/nh
Three Phase Power Analyser, without harmonics without graphics, 0.1% accuracy, nonvolatile memory to save 20 instrument settings.
   

Accessories

ACS1
Current clamp with connector to 4106A shunt input; 0-200A / 0-1000A, DC-1kHz, basic accuracy 2%. Other ranges and accuracies on request

ACS2
Portable printer (106 x 180 x 88mm) with Centronics interface and cable (400g)

ACS3
Soft carrying case for 4106A

ACS4
Set of test leads, max. 32A, 1.5m (2 red; 2 black)

ACS5
Shunt input connector

ACS6
Service Manual for 4106A

ACS7
Rack Mounting Kit

ACS8
Calibration Certificate traceable to National Standards

ACS9
USB - RS232 Adapter cable, USB 1.1 compatible, Windows 98/2000/ME/XP, > 1Mbps data transfer rate

 

Options
01
RS-232 Interface and Centronics printer output

02
RS-232 and IEEE-488 Interface, Centronics printer output

03
RS-232 and IEEE-488 Interface, Centronics printer output, 4 analog utputs, 8 analog inputs

03A
RS-232 and IEEE-488 Interface, Centronics printer output, 4 analog outputs with provision to also output total power, 8 analog inputs

04
3-Phase current sensor module 0-100A (supply by 4106A)

05
Operating software under DOS to control 4106A, read data and store data via RS-232 or IEEE-488 (National / Keithley interface)

06
4106A driver for Nat. Instrument LabView

07
Standalone 4106A software based on LabView

07M
Software for Motor Testing

07T
Software for Transformer Testing

08
TTL-input for external synchronisation

09
Network to form artificial neutral (mainly for frequency inverter measurements)

10
0-300A, 0-3000A flexible current clamps with connector to clamp input of Power Analyser (1 per phase). Basic accuracy: 1% at 50/60Hz

 

For more information about this product contact Tinsley

 