Single and Three Phase Power Analysers Model 4107A Series

Features

  • DC-300 kHz, 100 mA-50A, 1 V-1000 V
  • Large and bright display for up to 10 values
  • Line-to-line voltage, torque, slip, efficiency
  • Harmonics 1-63, IEC1000-3-2
  • High speed data logging for dynamic processes
  • High speed dynamic torque measurement
  • Advanced operating software under Windows
  • Accuracy grades 0.05, 0.1, 0.3 % (low cost)
  • Determines torque without using transducers

For more information about this product contact Tinsley

 

Single and Three phase Power Analysers Model 4107A Series

Introduction

The model 4107A High Speed Power Analyser is a very advanced measurement tool designed to offer the scientist, engineer, or technician more electrical and mechanical signal information than is normally found in one single instrument.

High performance - Easy to use

The Tinsley model 4107A Power Analysers are multipurpose instruments designed for laboratory and field use, for production testing and quality control. High accuracy, wide frequency range, and high common mode rejection are prerequisites for precision measurements on inverters, light ballasts, and ultrasonic transducers. The 4107A measurement capabilities and simplicity of use are unmatched.

Six menu soft keys M1 through M6 and five cursor soft keys control the instrument. The procedure to configure the display and interface, or to store instrument settings, such as the scaling factors, selecting the operating mode, selecting the input ranges, and many more features are self explanatory. You can save your personal instrument settings and have the unit start-up with your personal configuration at power-on.

Efficient data processing permits simultaneous measurements on all three phases of a three phase system and yields comprehensive signal information including rms and power values, harmonics, phase angles, and where applicable mechanical power output, torque, efficiency, and slip.

A programmable data logging function permits transfer of up to 500 rms and power values per second to a personal computer. The Tinsley operating software easily produces plots of power versus time or speed, or current versus time or frequency. The 4107A Power Analysers determines electrical motor torque applied to its load. No torque transducers are needed. Induction motors and dc motors require a speed transducer, synchronous motors do not. A special operating mode for dynamic torque measurement in the air gap of a generator or motor provides plots of torque versus speed, frequency or time, and helps to detect undesirable peaks.

Model 4107A Specifications

See the table below for a run-down of this product's specifications.

Voltage

7 ranges: 1V, 3V, 10V, 30V, 100V, 300V, 1000V

 

Frequency range

DC–300kHz

Coupling: AC/AC+DC

1Hz–300kHz / DC–300kHz

 

Crest Factor

3:1 at 50% full scale

Common Mode 50Hz / 100kHz

140dB/80dB

 

Built-in star point network

500kΩ Maximum Input: Hi to Lo / Lo to case

1000V / 600V

 

Accuracy 23°C ±3°C; rms, rdg= reading

 

1Hz–1kHz

±k(0.05% rdg + 0.07% range)

Accuracy Grades: k=1, k=2, k=6

 

1kHz–10kHz

±k(0.2% rdg + 0.2% range)

 

10kHz–80kHz

±k(0.04%/kHz rdg + 0.3% range)

 

80kHz–300kHz

±k(0.04%/kHz rdg + 0.3% range), typical

     
Current

10 ranges: 100mA, 300mA, 1A, 3A,10A; 1, 3, 10, 30, 100A

Clamp: 1A – 1000A

 

Frequency range: DC-300kHz

Coupling: AC/AC+DC

1Hz–300kHz / DC–300kHz

 

Crest Factor 3:1 at 50% full scale

Common Mode 50Hz / 100kHz

160dB/115dB

 

3A input: 3A cont./10A 5s; 50A input: 40A cont./50A 20s.

RI=30mΩ/ 3mΩ

 

Accuracy 23°C ±3°C; rdg= reading, rng= range

 

 

50A/Clamp input

3A input

Accuracy Grades: k=1, k=2, k=6

 

1Hz–500Hz

±k(0.07% rdg + 0.07% rng)

±k(0.07% rdg + 0.07% rng)

 

500Hz–2kHz

±k(0.5% rdg + 0.5% rng)

±k(0.8% rdg + 0.8% rng)

 

2kHz–10kHz

±k(0.8% rdg + 0.8% rng)

±k(2% rdg + 1% rng)*

 

10kHz–100kHz

±k(0.1%/kHz rdg + 0.8% rng)*

±k(0.1%/kHz rdg + 2% rng)*

 

Lowest range multiply percentage figures by 2. * typical

 
   
Power

70 Ranges corresponding to the product of the voltage ranges x current ranges.

DC-300kHz

 

Accuracy 23°C ±3°C; 50A Clamp input, 3A input

 

 

1Hz–1kHz

Add accuracy percentage figures of current and voltage input

PF = 0 to ±1

 

1kHz–100kHz

Add accuracy percentage figures of current and voltage input,
add ±1% (1-PF)/kHz of range

 

 
 
Computed Values

Reactive Power, Var = ±(VA2 - W2)1/2;
Apparent Power: VA = Arms Vrms;

 

Power Factor: PF = W/VA; Crest Factor:
CF = Ap/Arms, Vp/Vrms; Maximum = Ap, Vp;

Add accuracy percentage figures of values involvedin computation.

 

Minimum: -Ap, -Vp; PtP: Maximum–Minimum; Impedance: Z = Vrms/Arms

 

Total Harm Dist:
THD1 = (Irms2 - IFund2)1/2/Irms
THD2 = (I2rms2 + I3 rms2 + Irms2)1/2/Irms

     
Mechanical Values

Total input power, Nm/s; Output power to load, Nm/s; Torque at axis of rotating machine,Nm; Speed, rpm; (for induction and dc machines an external frequency input via analog input 0 is required); Efficiency: Eta = Output power / Input power; Slip: Slip=(fi - fo)/fi

Motor version including analog input, RS-232, and software.

Dynamic
Torque

Dynamic torque in air gap of rotating machines versus time or speed; moment of inertia of rotor.

 
   
Harmonic Analysis

Frequency range of fundamental 4Hz–50kHz

Harmonic 1–63

 

Accuracy: harmonic current and voltage, same as rms current and rms voltage

Add accuracy percentage figures of values involved in computation.

 

Computed values: harmonic power, harmonic phase angle (power factor); harmonic impedance.

   
Frequency

2Hz-100kHz; A; V, or external triggered: ±0.1%

   
Integrator

Energy, Accuracy: Wh, VAh; Basic accuracy of integrated quantity.

     
Data Logging

Output values, speed and duration is programmable; maximum speed: 10 values from a 3-phase system in 20ms.

Range of signal frequency 5Hz to 1kHz

   
Current harmonics

Current harmonics in a 3-phase system are determined according to IEC1000-3-2

     
Display

Blue liquid crystal graphic display with FL backlight

64x120mm;128x240 pixels

   
Power

AC, 50–60Hz; Fuse; 0.2AF, Power 115/230V; 15VA

     
Dielectric Strength

Current inputs to case; Voltage inputs to case; Line input to case

3kV; 2kV; 3kV 50Hz/1 min

     
Dimensions

H x W x D; weight

150x235x320mm; 4kg

   
Options

RS232 interface

 

10 High speed analog inputs (1 frequency, 9 dc voltage)

fx, ±5V

 

10 Analog outputs

±5V

 

Windows operating software 95,98,NT,ME,2000,XP

 

Versions

4107A-1/0.3L
Economy Single Phase High Speed Power Analyser, 0.3% accuracy, saves 12 instrument settings, no data logging, no IEC1000-3-2.
4107A-1/0.3
Single Phase High Speed Power Analyser, 0.3% accuracy, saves 12 instrument settings, includes data logging, no IEC1000-3-2.
4107A-1/0.1
Single Phase High Speed Power Analyser, 0.1% accuracy, saves 12 instrument settings, includes data logging and IEC1000-3-2.
4107A-1/0.05
Single Phase High Speed Power Analyser, 0.05% accuracy, saves 12 instrument settings, includes data logging and IEC1000-3-2.
4107A-3/0.3L
Economy Three Phase High Speed Power Analyser, 0.3% accuracy, saves 12 instrument settings, no line-to-line measurement, no data logging, no IEC1000-3-2.
4107A-3/0.3
Three Phase High Speed Power Analyser, 0.3% accuracy, saves 12 instrument settings, includes line-to-line measurement and data logging, no IEC1000-3-2.
4107A-3/0.1
Three Phase High Speed Power Analyser, 0.1% accuracy, saves 12 instrument settings, includes line-to-line measurement, data logging, and IEC1000-3-2.
4107A-3/0.05
Three Phase High Speed Power Analyser, 0.05% accuracy, saves 12 instrument settings, includes line-to-line measurement, data logging, and IEC1000-3-2.
4107A-3/0.05M
Three Phase High Speed Power Analyser for dynamic motor testing torque transducer required), 0.05% accuracy, saves 12 instrument settings, includes line-to-line measurement, dynamic torque easurement, EC1000-3-2, RS-232 interface, frequency/analog in-out, data logging and operating software for motor testing.
4107A-3/0.05T
Three Phase High Speed Power Analyser for transformer testing according to IEC60076-1, 0.05% accuracy, saves 12 instrument settings, includes line-to-line measurement, data logging, IEC1000-3-2, RS-232 interface, and operating software.

 

Accessories

ACS1
Current clamp with connector to 4107A clamp input; 0-200A
/ 0-1000A, DC-1kHz, 2%
ACS2
Current clamp with connector to 4107A clamp input; 0-
200A, DC-20kHz, 1%
ACS3
Soft carrying case for 4107A
ACS4
Set of test leads, max. 32A, 1.5m (2 red; 2 black)
ACS5
Clamp input connector
ACS6
Service Manual 4107A
ACS7
Rack Mounting Kit
ACS8
Calibration Certificate traceable to national standards
ACS9
USB - RS232 Adapter cable, USB 1.1 compatible, Windows
98/2000/ME/XP, > 1Mbps data transfer rate

 

Options

01
High Speed RS232 Interface, external synchronization input, Windows Operating Software (95,98,NT,2000,ME,XP) without data logging capabilities, without IEC1000-3-2
02
High Speed RS232 Interface, external synchronization input, trigger input, Windows Operating Software (95,98,NT,2000,ME,XP) including software for high speed data logging and IEC1000-3-2 testing
03
One frequency input with selectable frequency range, nine analog inputs (±5V / ±10V), ten analog outputs(±5V)
04
0-100A Three Phase Current Sensor Module including supply
05
0-300A, 0-3000A flexible current clamps with connector to clamp input of 4107A (1 per phase). Basic accuracy: 1% 50/60 Hz

 

For more information about this product contact Tinsley

 