Resistance Thermometry
Helium Filled Platinum Sheath Resistance Thermometer Model 5187L
Specifications |
- Temperature range: 13.8K to 303K (M.P. of Gallium)
- Resistance at 0 °C: R(tp) 25 Ohm +/- 0.5 Ohm
- Reproducibility: +/- 1mK
- Accuracy: +/- 2mK
- R(Ga)/R(tp): Ratio not less than 1.11807
- Response time: 2 seconds
- Gas Filling: Helium at 1/3 atmosphere at 20 °C
- Calibration: NPL Calibration available
- Dimensions: Length 65mm, Sheath length 47mm,
Sheath Diameter 5.2mm
|
Shipping Specification |
- Case size: 160 x 55 x 120 mm
- Net weight: including thermometer 268g
Suitable for use with Tinsley type 5840D Senator Bridge
For more information about this product contact Tinsley
|
|

|
This Precision Platinum Resistance Thermometer is constructed in a 5.2 mm platinum sheath 47 mm long with platinum leads passing through a glass seal making the overall length approximately 65 mm.
Construction
The Resistance Temperature Dependent (RTD) element is constructed from a high purity platinum wire and is formed and processed to produce a strain free construction. This is then introduced into a platinum sheath which is helium filled to approximately 1/3 atmosphere at 20°C to reduce the response time to approximately 2 seconds.
Application
The precision thermometer is suitable for use in the temperature range 13.08K to 330K and may be used totally immersed in a liquid, provided the liquid is a good electrical insulator.
For more information about this product contact Tinsley
|