Resistance Thermometry

High Precision Platinum Resistance Thermometer Type 5187SA

Specifications

  • Temperature Range:  -189 °C to 660 °C
  • Calibration: Suitable for calibration to ITS-90 up to max temp 660.323 °C (F.P.Aluminium)
  • Resistance at 0°C (Rtp): 25 Ohm +/- 0.5 Ohm
  • R(Ga)/R(tp):  Ratio not less than 1.11807
  • R(Me)/R(tp):   Ratio not greater than 0.844235
  • Reproducibility: +/-0.001 °C
  • Gas Filling:  Dry air at 1/3 atmosphere
  • Basic Accuracy:  +/-0.001 ° C
  • Dimensions: Sheath diameter 6.5mm to 7.5mm length 480mm. Head diameter 23mm length 60mm
    Overall length 540mm 
    Immersion depth 300mm
  • Self heating: 0.002 °C to 0.003 °C with 1mA and the thermometer bulb immersed in unstirred water
  • Leads: Supplied with 4 metres of PTFE insulated cable fitted with gold plated copper spade tags
  • Case Size: 67 x 16 x 7 cm
  • Net Weight: Including thermometer 1.15kg

For more information about this product contact Tinsley

 

High Precision Platinum Resistance Thermometer Type 5187SA

This four terminal Precision Platinum Resistance Thermometer is designed to realise, with the highest accuracy, the International Temperature Scale ITS-90 over the range -189°C to 66O°C. The design is the result of many years practical experience and cooperation with the National Physical Laboratory, Teddington, where primary resistance and other standards are maintained.

Construction

The Resistance Temperature Dependent (RTD) element is of pure platinum, coiled and mounted in a strain free construction. The former is of fused silica and great care is taken to ensure freedom from contamination. All the joints are welded, the four leads from the element to the seal in the thermometer head are heavier gauge platinum thereby avoiding the generation of thermal e.m.f.'s at the junction with the element. The leads are brought through a hermetic seal at the head of the thermometer and joined via low loss terminals to copper flex in a specially constructed cable with four conductors. The cable is made with PTFE insulation to ensure low dielectric loss so that the thermometer may be used with either d.c. or a.c. measuring systems. The terminations are gold plated copper spade tags. The thermometer tube is specially treated to avoid radiation loss by the piping effect in its walls. Each thermometer is supplied with a purpose-made carrying case.

Stainless Steel Sheath

To reduce the possibility of damage, the type 5187SA thermometer may be supplied with a protective stainless steel sheath, outside diameter 8-9mm, fitted into the thermometer head by means of three screws. The thermometer is calibrated without the sheath and when this is in position the response time is increased to about 20 seconds.

Calibration

The table below sets out the types of calibration available from NPL and their uncertainties.

Calibration of platinum resistance thermometers to international temperature scale
Point
Temperature °C
Uncertainty ° C
 
 
Range 1
Range 2
Range 3
Range 4
TP argon
-189.3442
 
0.002
0.002
0.002
TP mercury
-38.8344
0.0005
0.0005
0.001
0.002
TP water
0.01
0.0005
0.0005
0.001
0.002
MP Gallium
29.7646
0.0005
 
 
 
FP indium
156.5985
 
0.001
 
 
FP tin
231.928
 
0.001
0.001
0.002
FP zinc
419.527
 
 
0.001
0.002
FP aluminium
660.323
 
 
 
0.004
           

The type 5187SA Platinum Resistance Thermometer's maximum calibration will be 660.232°C.

Note: TP - Triple point, MP  - Melting point, FP - Freezing point.