Tinsley Precision Instruments
Tinsley Precision Instruments

page location: home > products > strain measuring equipment > model 8691

Strain Measuring Equipment

Single Channel Strain Meter Model 8691

Single Channel Strain Meter Model 8691The Model 8691 single channel direct reading digital instrument is useful in a wide range of applications for accurate measurement of strain with resistance strain gauges in ¼, ½ or full bridge configurations. Bridge Voltages and Mode are selected by push button. The strain measurement instrument is housed in a two tone case incorporating adjustable carrying handles which also acts as a means of inclining the instrument for easier bench top use.

Battery operation option makes it useful as a portable strain meter.

For more information about this product contact Tinsley

Specifications

See the table below for a run-down of this product's specifications.

Channels
Single
Range
± 19999 microstrain
Resolution
1 microstrain
Linearity
0.02% of Full scale
Gauge Resistance
¼ bridge: - 120 Ω or 350 Ω
½ bridge: - 80 Ω upward
Full bridge: - 120 Ω upward
Gauge Factor Range
1 to 2.99 in 0.01 steps
Bridge Voltage
2V, 5V or 10V DC
Bridge Modes
¼ , ½   or Full bridge
Bridge Zero
Coarse adjustment is in 9 overlapping steps up to 10,000 µ strain. ± 2% resistance imbalance is covered
Fine adjustment is by a locking 10 turn front panel control with a range of 700 µ strain
Zero Drift
less than 0.5 microstrain / °C
Gain Drift
less than 0.005% / °C
Input Impedance
1000 MΩ
Gauge Connections
Screw terminals are at the rear and provide for ¼ , ½   or full bridge connections
Analog Output CRO
± 2V. Equivalent to ± 20,000 microstrain
Frequency Response
DC to 20 KHz
Power
Mains 115/240V AC. 50/60Hz. Slide switch at rear. BATTERY (optional). 2 x NiCad rechargeable batteries built-in, 8.4V, 1.2AH. Full charge of 40 hours approx.
Dimensions
300 x 110 x 450mm
Weight
5.0kg
   